Measurement principle
The sensor behaves pretty much like a potentiometer; since resistance is realized by a liquid electrolyte, sensor requires power in AC. The supply voltage normally used varies from 0.5V to 5V, with a frequency range from 20 Hz to 20,000Hz.
The instrument is provided with a specific signal conditioner for the DC power supply and for adjusting the signal; the average life is estimated at 400.000 hours.
The sensor is supported by a suitable base which also monitors and calibrates the sensor, thanks to a device specifically designed and made by us with high quality precision mechanics.
The base which supports the sensor can perform small rotations, operated manually, by a mechanical device fixed to it. On the same base, there are three special feet that support a high accuracy bubble level clinometer (accuracy 2” on a range of ± 30 ‘).
When the horizontal clinometer sensor is placed on the base, a measurement can be taken using either the automatic sensor or the manual clinometer; the high accuracy of the latter instrument allows verification of the zero value of the sensor (as measured at the time of installation).
When the base is rotated, it measures the extent of the tilt both with the bubble level clinometer and with the electrolytic sensor. The eventual difference in the measurements gives us the parameters for the correction of instrumental drift; correction which is carried out by acting on the electronic gain of the sensor conditioning.
The base’s mechanical rotation device ensures the perfect repositioning of the base in its initial position (before the test). Positioning accuracy guarantees an margin of error ten times smaller than instrumental error.
This method allows the elimination of the most important causes of error in these types of instruments
Conditioner and converter guarantee:
- Conditioner operating temperature: 25 ° C to + 70 ° C
- Temperature drift of the converter 4-20mA: 100ppm / ° C
- Supply Voltage: 24Vdc-25% / + 50%
Optionally, an 3000VDC electronic buffer is available for the galvanic insulation of the supply voltage and signal circuits.
Technical Specifications
Full Range |
60’ |
Measurement Error
In range 9’
In range 9’÷ -30’ and +9’ ÷ +’30 |
<0,15% f.s.
<4% f.s. |
Maximum Absorption |
<150 mA |
4-20mA Converter |
13 Bit |
Coefficient temperature zero point (null) |
5” / 50 °C |
Coefficient temperature scale factor |
0,06% f.s./°C |
Time Coefficient |
TC = 2,5 sec |
Working temperature range |
From 25 °C to +70 °C |
Supply Voltage |
24 Vdc +25% / +50% |
Optional |
Electronic Buffer 3000 Vdc for the galvanic isolation between supply voltage and signal circuitry |
Accessories and related products
Junction Box |
Available into different models for the connection of different instruments |
Selection and Measurement Box |
Complete measurement box with MUX cards for the automatic selection |
Measurement Box |
Simple measurement box |
Cable |
Multipolar Cable |
DEC3000 |
Portable Datalogger |
CUM3000 |
Multichannel Datalogger |
MUX |
Multiplexer for the connection of different sensors to the Datalogger |
Bubble Level Clinometer |
For the comparison between the electric measurement and correction of drifts. |
Technical Assistance
If you have any requests or questions about our instruments or if you have special needs that require different solutions from the standard, please contact us. Our team will provide all the necessary information and will be very happy to work with you to study, develop and customize instruments and solutions suitable for your specific needs.
All information in our datasheet is subject to change without prior notice.
Please check the datasheet/details carefully and contact Pizzi Instruments for further information